![]() ![]() Ninety-three percent of the Glucocard Memory 2 and 96% of the Accutrend sensor results were within 15% of the comparison method results. Ninety-eight percent of both Glucocard Memory 2 and Accutrend sensor results were within 20% of the comparison method values. Results of both blood glucose meters correlated closely with the hexokinase/glucose-6-phosphate dehydrogenase laboratory method. Results were linear over a range of 1.6 mmol/l -29.7 mmol/l for Glucocard Memory 2 and 1.6 mmol/l -33.3 mmol/l for Accutrend sensor. ![]() Between-day coefficients of variation for Glucocard Memory 2 were 4.8% and 3.5% at glucose concentrations of 3.9 mmol/l and 17.2 mmol/l, respectively and for Accutrend sensor they were 3.8% and 2.9% at glucose concentrations of 3.8 mmol/l and 18.7 mmol/l, respectively. ![]() Within-run coefficients of variation for Glucocard Memory 2 were 6.3%, 3.9% and 2.4% at glucose concentrations of 1.7 mmol/l, 5.8 mmol/l and 11.7 mmol/l, respectively: for Accutrend sensor these were 15.2%, 5.0% and 1.2% at respective concentrations of 0.9 mmol/l, 4.2 mmol/l and 19.6 mmol/l. In the hypoglycaemic concentration range, the imprecision of Accutrend sensor was much higher than recommended by the American Diabetes Association. Both glucose meters produced acceptably precise results in the hyper- and normoglycaemic concentration ranges. The performance and practicability of 2 blood glucose meters (Glucocard Memory 2 and Accutrend sensor) were evaluated. ![]()